Short description
The ARES rheometer is a true strain-controlled instrument, where the application of strain and the measurement of stress is truly separated. The bottom tool is rotated by a motor, while the force rebalance transducer measures the torque response on the top tool. While not as sensitive as some modern stress-controlled rheometers, this machine still has many advantages:- Less influence of tool inertia on oscillatory measurements at high frequency.
- Results in steady shear do not depend on tool angle, no problems with angle-dependent friction of air bearings.
- Accurate stress-relaxation tests.
Things to remember
-
Before handling the tool:
- turn off motor
- set to transducer 2
- Insert lower tool only if bath and tool are at the same temperature. (thermal expansion .. :-)
- Do not change motor to dynamic mode with sample loaded.
-
- (if absolutely necessary, you can turn off the motor, turn the lower tool to the "zero" position and then change to dynamic mode)
- Do not use "Set Gap" in the Gap control panel before running "Zero Gap".
- Do not leave the transducer in overload mode (click the yellow warning button)
Using the ARES
- Get training
- Reserve it
- If anything unusual happens during use, please ask
one of the people responsible for the rheometers. If the
main air pressure drops below 70 psi,
stop all measurements, do not touch the tools and contact
one of the following emergency contacts:
- Harvard facilities office: (617) 495-2840
- Harvard Police (during the night): (617) 495-1212
Measuring Geometries
Geometry |
Dimensions |
Sample Volume (approx.) |
Notes |
|
Parallel plate 50 mm
|
|
2 mL with 1 mm gap | Shear gradient exists across the sample. |
|
Cone/Plate 50 mm
|
|
2 mL | Constant shear profile. |
|
Couette 34 mm
|
|
13.5 mL |
Constant shear profile.
Large volume required. |
|
Couette 17 mm
|
|
3 mL | Constant shear profile. |
|
Double wall couette
|
|
6.5 mL |
Constant shear profile.
Home built tool. Heavy.. sometimes problems with normal force overload. |
- Transducer 1:
- Low torque limit: 4e-7 Nm
- High torque limit: 1e-3 Nm
- Transducer 2:
- Low torque limit: 4e-6 Nm
- High torque limit: 1e-2 Nm
- Please use transducer 1 only if you need extra sensitivity at low torques. Always switch to transducer 2 whenever you are handling the upper tool. Remember that too much force or torque on the upper tool can damage the transducers.
Other information
Technical support
- Patel Scientific is a company that specializes in servicing ARES rheometers, their support number is (908) 510-4407.
- For support from TA Instruments, call (302) 427-4000.